Turck
Turck
Temperature Detection – With Current or Voltage Output and PNP/NPN Transistor Switching Output
SKU: 100004382 Type: TS700-L050-16-LI2UPN8-H1141
Available 3–5 business days delivery
€330.20 / pieces
€392.94 incl. VAT
Features
- • Screw-in adapter with process connection NPT1/2" male thread included in delivery
- • 4-digit, 2-colored, 12-segment display, rotatable by 180°
- • Upper part of housing can be rotated up to 340°
- • Materials in contact with media: 1.4404 (316L), CR O-ring
- • Immersion length: 50 mm
- • 17…33 VDC
- • NO/NC contact, PNP/NPN output, analog output (current/voltage), IO-Link
- • Various IO-Link mapping profiles can be selected
Functional principle
The TS+ product series is a compact processing unit with a 4-digit, 12-segment display that can be freely rotated by 340°. Compact units with (TS700) and without (TS720) temperature sensors as well as different output variants are available.
Tests/approvals
| C | Green On Temperature in °C |
| FLT | Red On Error displayed |
| FLT | (For error pattern in combination with LEDs see manual) |
| FLT | Off No errors displayed |
| °FKΩ | Green On Temperature in °FGreen On Temperature in KGreen On Resistance in Ω |
| I and II (switching point of LEDs) | Yellow On Switching output |
| I and II (switching point of LEDs) | -NO: switching point exceeded/within window (active output) |
| I and II (switching point of LEDs) | -NC: switching point undershot/outside window (active output) |
| I and II (switching point of LEDs) | Off Switching output |
| I and II (switching point of LEDs) | -NO: switching point undershot/outside window (inactive output) |
| I and II (switching point of LEDs) | -NC: switching point exceeded/within window (inactive output) |
| Status LED | s |
| IO-Link proc | ess data image |
| Included in delivery | Screw-in adaptor with process connection |
| Included in delivery | 1/2" NPT male thread |
| LOC | Yellow On Device locked |
| LOC | Flashing "Lock/unlock" process is active |
| LOC | Off Device unlocked |
| Byte n | 14 Bit Process Value (TEMP) State Out 2 State Out 1 |
| UL registration number | E516036 |
| For detailed | description of the display patterns and flashing codes, see manual |
| Approvals | cULus |
| PWR | Green On Operating voltage applied |
| PWR | Device is operational |
| PWR | Flashing Operating voltage applied |
| PWR | IO-Link communication active |
| PWR | (inverted flash with T on 900 ms and T off 100 ms) |
| LED | Color Status Description |
| Bit | 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 |
Display
| Display switch state | 2 × LEDs, Yellow |
| Unit display | 4 x green LED (°C, °F, K, Ohm) |
| Display | 4-digit, two-colored (green/red), 12-segment display, rotatable by 180° |
Reference conditions acc. to IEC 61298-1
| Auxiliary power | 24 VDC |
| Humidity | from 10 to 95 % |
| Atmospheric pressure | from 860 to 1060 hPa |
| Temperatur (-) | 15…+25 °C |
IO-Link
| Transmission rate | COM 2 (38.4 kBaud) |
| Vibration resistance | 20 g (10…3000Hz), |
| Transmission physics | corresponds to 3-wire physics (PHY2) |
| Elektrischer Anschluss | Connector, M12 × 1 |
| Frame type | 2.2 |
| Ambient temperature | from -40 to 80 °C |
| Process connection | 1/2'' NPT male thread |
| Included in the SIDI GSDML | Yes |
| Storage temperature | from -40 to 85 °C |
| IO-Link specification | V 1.1 |
| Housing material | Stainless-steel/Plastic, 1.4404 (AISI |
| Housing material | 316L)/Grilamid TR90 UV |
| Shock resistance | 50 g (11 ms) , DIN EN 60068-2-27 |
| Accuracy | ± 0.1 K |
| Parameterization | FDT/DTM/selection of various mapping |
| Parameterization | profiles |
Temperature range
| Short-Circuit/Reverse Polarity Protection | yes / yes |
| Pressure resistance | 300 bar |
| Immersion depth (L) | 50 mm |
| Protective measure | SELV, PELV according to DIN EN 61140 |
| Voltage drop at Ie | ≤ 2 V |
| Protection type and class | IP66 |
| Protection type and class | IP67 |
| Protection type and class | IP69K / III |
| Accuracy | 0.15 °C + 0.002 ∙|t| (-30…300 °C) |
| Measuring range | -50…150 °C |
| Measuring range | -58…302 °F |
| Operating voltage | from 17 to 33 V |
| Electrical safety class | III |
| Messelement (nach DIN IEC 751) | Pt-1000 probe, DIN EN 60751, class A |
| Ansprechzeit (bei ruhender Luft) | t0.5 = 3.5 s/t0.9 = 9.5 s in water at 0.2 m/ |
| Ansprechzeit (bei ruhender Luft) | s |
Analog output
| Current output note | Each invertible |
| Current output | 4…20 mA |
| Voltage output note | Each invertible |
| Alternative voltage output | 0… 5 V |
| Alternative voltage output | 1… 6 V |
| Alternative voltage output | 0.5…4.5 V |
| Repeatability | 0.1 K |
| Note | 0.1% of full scale applies to temperatures |
| Note | > 200 °C |
| Alternative current output | 0…20 mA |
| Alternative current output | 20…0 mA |
| Accuracy | ± 0.3 K |
| Voltage output | from 0 to 10 V |
Programming
| Vibration resistance | 20 g (10…3000Hz), |
| Ambient temperature | from -40 to 80 °C |
| Storage temperature | from -40 to 85 °C |
| Programming options | Switching logic (PNP, NPN, Auto*), |
| Programming options | switching/release point; hysteresis/window mode; NO/NC; unit; IO-Link mapping |
| Programming options | profile; analog output (U, I, Auto*); 0 (4)… |
| Programming options | 20 mA, 0…(5) 10 VDC, 1…6 VDC, 0.5… |
| Programming options | 4.5 VDC; * factory settings |
| Degree of protection (IP) | IP67 |
| Degree of protection (IP) | IP66 |
| Degree of protection (IP) | IP69K |
Switching outputs
| Switching point accuracy | ± 0.2 K |
| Release position | -50…+149.8 °C |
| Switching point | -49.8…+150 °C |
| Switching cycles | ≥ 100 mil. |
| Output function | NO/NC programmable, PNP/NPN |
| Switching point distance | ≥ 0.2 K |
| Communication protocol | IO-Link |
| Switching current | up to 0.25 A |
| Rated operating current | 0.25 A |
Output
| Output 1 | Switching output or IO-Link mode |
| Output 2 | Analog or switching output |